Input Power: Standard | 9 Vdc +/- 0.5Vdc @ 5VA (depends on external loads) provided by external AC wall transformer (provided)., non-isolated, 100-240Vac 50/60Hz | Input impedance: | >100k Ohms |
Option: | Isolated 12-24 Vdc input power available (not compatible with internal battery pack option below). | Measurement Rate: | Up to 100 samples/sec per channel. |
Option: | Internal battery pack provides uninterrupted operation and controlled shutdown during blackout. 6Vdc, 2400mAH NiMH - Backup Time: 8 hours typical (depends on external load). | Math Function: | Y=mx + b; average, hi peak, low peak. |
DC Sensor Power Output: | 2 outputs 5Vdc @ 50mA to power external sensors. | Media: | Compact Flash to 2GB |
Number of Channels: | 2 universal, user selectable. | Display: | LCD Graphics, 160 x 80 pixels, black FSTN with white LED backlight. User controlled backlight level and contrast adjust (electronic). |
Isolation: | 300V AC/DC channel input to chassis ground | Display Modes: | Trending (horizontal), Large dual digital readout, mixed mode. |
DC Input: | Voltage Ranges: 0-250mV; 0-1.25V; 0-2.5V, 0-5V; 0-12.5V; 0-25V | User Interface: | 5 button keypad (dual function buttons). |
| Accuracy: 0.1% of reading Resolution: 0.025% of full scale | Clock: | Auto leapyear and daylight savings adjustment. Internal battery backup. |
| Current Ranges: 0-20mA; 4-20mA; 0-50mA; 10-50mA | Relay Output: | Two alarm outputs: 30V 0.5A Form A relays. |
| Accuracy: 0.1% of reading excluding 250 ohm external shunt (required) | Opto-isolated Input: | One input, 5 to 12Vc activation @ 10mA typical. |
| Resolution: 0-25% of full scale. | Audible: | Internal beeper (multiple tones). |
Thermocouple: | Accuracy: 0.3% of full scale (typical). Ambient Temperature Sensor Accuracy: +/-1.5°C (see datasheet for full specs.) | Dimensions: | Front panel: 96mm (1/4 DIN) x 152mm (3.78 x 3.78 x 6 inches). |
RTD: | Accuracy: 0.3% of full scale (typical) Resolution: 0.1°C Internal Current source: 1mA (see datasheet for full specs) | Environmental: | Indoor Use Only Installation Category II per IEC 664 Pollution Degree Level II per IEC61010-1 Temperature: -10°C to 50°C operating per IEC 61010-1 Humidity: Maximum relative humidity 80% for temperature up to 31°C, decreasing linearly to 50% relative to humidity at 40°C. |
Frequency Input (Hz) Range: | 0-10,000 Hz/ 0-600,000 RPM | Optional: | USB: Front panel USB 2.0 slave port for data and remote real-time display when using Navigator software. |
Speed Input (RPM): | Accuracy: Freq: +/-1 Hz; RPM +/1 RPM below 9,999 RPM Input: Low <1.0Vdc; High >3.0 <12.0Vdc Pulse Width: 10 microsecond minimum. | | Ethernet: Real time data display, historic data transfer, remote control and recorder configuration when using Navigator Software. Built in web server displays current measured values. |